GDCY-20 MPULSE TEST SYSTEM

 

I. Application
 
 
This impulse test system is used for testing SPD of grade III.
 
 
II. Design
 
Based on IEC61000-4-5 and GB/T17626.5-1999, which stipulate voltage waveform, characteristic and performance of signal generator, and coupling/decoupling network.
 
Based on IEC61643-11 and GB/T18802.1-2002, which stipulate testing of SPD of grade I and II, and composite wave testing of grade III.
 
 
 
III. Main Specifications
 
 
1. Voltage waveform and output voltage range
 
Open voltage Uac: wave front time 1.2μS±30%, half peak time 50μS±20%, peak value
 
±3%;
 
In the peak of wave, the voltage ≥5% of peak value over-shoot or oscillating is not permitted. Open voltage peak value 1kV-20kV.
 
2. Charging voltage displayed by pointer voltmeter and digital voltmeter. Forward direction or reverse direction impulse is realized.
 
3. Display impulse voltage, accuracy is ±1%.


 
 
 
 
 
 
 
4. Output voltage measurement
 
Impulse voltage divider will be supplied.The accuracy of voltage ratio is ±1%. It output to BNC port of the panel, for oscillograph connection by user.
 
5. Enough safety and security device, to ensure safe operation.
 
6. All the components are equipped in the cabinet, easy and safe to operate.
 
7. Rated parameter
 
Input power: 220VAC, 5kVA
 
Rated charging voltage: 20kV
 
Total capacitance: 9μF*1
 
Rated energy: 18kJ.
 
8. Voltage waveform and amplitude
 
When charging voltage is 20kV, the generator can output 1.2/50μs waveform of impulse voltage, voltage amplitude 20kV.
 
Duty cycle: Continuous working at rated voltage and every 120s charging and discharging once.
 
Continuous working at 2/3 of rated voltage and every 10-60s charging and discharging once.
 
9. The composition of generator
 
HV DC charging part
 
Energy storage and discharge part of main body
 
Wave-adjustable inductors and resistors, and other adjustable components
 
The ignition sphere gap is cooper ball with tungsten inlay. The control mode is mechanic and impulse way, which can endure big current.
 
 
 
VI. System composition


 
 
 
 
 
 
 

Name
 
Basic parameter
Qty
Remark
 
 
 
 
 
 
 
 
1
Main body
Total
 
capacitance
9μF,
1
Standard
 
 
 
output voltage 20kV
 
 
 
 
 
 
 
 
 
2
Measurement
Voltage divider 20kV
 
1
Standard
 
device
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
3
Control system
PLC
control,
touch
screen,
1
Standard
 
 
 
isolated
power
supply,
 
 
 
 
 
safety interlock
 
 
 
 
 
 
 
 
 
 
4
Digital
 
TDS2012C model
 
1
Optional
 
oscillograph
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
5
Measurement
PC,
with
LCD
screen,
1
Standard
 
and
analysis
automatic reading waveform
 
 
 
system
 
parameters,
can
be
 
 
 
 
 
connected with oscillograph,
 
 
 
 
 
storage data and waveform,
 
 
 
 
 
print result etc.